Skip to content

Oct 13, 2011 - Masimo Launches 2011 Radical-7 with Rainbow Acoustic Monitoring

Masimo Launches 2011 Radical-7 with Rainbow Acoustic Monitoring and In Vivo Adjustment

IRVINE, Calif., Oct. 13, 2011 /PRNewswire/ -- Masimo (NASDAQ: MASI) announced FDA 510(k) clearance and CE Mark of its new Radical-7® with rainbow® Acoustic Monitoring capability for continuous display of the acoustic respiration rate (RRa(TM)) waveform and measurements. Also new to the Radical-7 is an In Vivo Adjustment(TM) (pending FDA 510(k) clearance in the U.S.) that allows clinicians to, for the first time, adjust the noninvasive measurements to the specific patient and laboratory reference device they use for invasive blood testing.

The new Radical-7 provides clinicians with the option to measure RRa and display the acoustic waveform directly on the monitor screen with the pleth waveform overlayed--allowing them to observe changes in breathing upon inhalation and exhalation. With this new display capability, clinicians may be able to more readily detect respiratory pause events where there is an absence of breathing, high ambient noise that can degrade the acoustic signal, and improper sensor placement.

Respiration rate is defined as the frequency of breathing expressed as the number of breaths per minute and is considered a critical vital sign in assessing the physiological status of hospitalized patients; however, current methods for respiration rate monitoring are limited by reliability or patient tolerance. In contrast, Masimo rainbow Acoustic Monitoring is completely noninvasive and virtually unnoticeable to the patient--featuring an innovative adhesive sensor with an integrated acoustic transducer that is easily and comfortably applied to the patient's neck to detect upper airway acoustic vibrations on the surface of the skin during the respiratory cycle. Using patented acoustic signal processing that leverages Masimo's revolutionary Signal Extraction Technology (SET), the respiratory signal is separated and processed to display continuous RRa measurements.

In Vivo Adjustment Allows Users to Account for Individual Patient Bias from Laboratory Reference Device

The measurement of blood components such as oxygen saturation, methemoglobin, carboxyhemoglobin, and total hemoglobin have inherent and expected variability within and between noninvasive and invasive measurement techniques and from patient to patient. For example, all of the noninvasive measurements use a technique coined 'patient calibration' which takes the noninvasive monitor's detected light that is transmitted or reflected from the patient at different values and matches them to empirically gathered invasive blood measurements from a small number of volunteers and in some cases patients. However, individuals may vary from the pool of patients used in the 'patient calibration' and hence provide a bias for the particular patient. In addition, many clinicians are unaware that laboratory reference devices that measure oxygen saturation or hemoglobin can have great variability from each other. In fact, hemoglobin can vary 2 g/dL or more from one manufacturer's product to another. And, physiologic factors such as the blood source (venous or arterial), site and time of blood draws, blood draw technique, and patient body position are recognized in the clinical literature to add variability to hemoglobin levels. As a result, noninvasive and invasive measurements can vary in the same patient depending on the methods and invasive equipment used.

The In Vivo Adjustment feature available in the new Radical-7 now provides the option to adjust the noninvasive oxygen saturation (SpO2), total hemoglobin (SpHb®), carboxyhemoglobin (SpCO®), and methemoglobin (SpMet®) values displayed on the monitor to the laboratory reference value used at the site by patient. With this new feature, clinicians can adjust the noninvasive value at the beginning of a monitoring period to account for individual patient variation and the laboratory reference value and continuously trend the noninvasive value to the patient and reference laboratory value until the next blood sample and laboratory analysis.

"With the addition of the new In Vivo Adjustment feature, Masimo refines its solution for three major limitations that have always plagued pulse oximetry: 1) the inability to measure-through motion and low perfusion (virtually eliminated by Masimo SET); 2) the inaccuracy of SpO2 in the presence of dyshemoglobins (they can now be monitored with SpCO and SpMet); and 3) the inability of one calibration curve to fit all patients (an individual patient's SpO2 can be adjusted to a reference arterial blood gas measurement)," stated Steven Barker, PhD, MD, Chairman of Masimo's Scientific Advisory Board.

Juan Soliveres, M.D., University Hospital in Valencia, Spain, stated, "There are many invasive devices that measure Hb and the results they report may be different from one device to the next. Even laboratory co-oximeters will report different Hb values when the same blood sample is run through different machines of the same make and model. When comparing invasive Hb measurements with noninvasive SpHb measurements, an In Vivo Adjustment feature--to remove measurement bias and calibrate SpHb measurement relative to the lab device--is an important enhancement that should contribute to improved Hb correlations between these measurements."

In addition, the new Radical-7 now features a larger number display with pulse oximetry measurements that are 40% larger than before to improve visibility. And for added safety, the device now includes an optional Alarms All Mute feature that allows the customer to enable or disable the ability for users to mute (silence) alarms on the monitor and adds an additional layer of protection that password-protects this feature--enabling only qualified clinicians to turn-on the Alarms All Mute feature. This feature safeguards against indiscriminate and accidental silencing of monitor alarms that can place patients at increased risk of unrecognized distress and deterioration, while enabling qualified clinicians (through password access) the ability to focus their attention on immediate patient care, interventions, and intensive care resuscitations in sensitive care situations without the audible distractions of monitor alarms.

Masimo Founder and CEO, Joe Kiani, stated, "With the latest Radical-7, we have once again expanded what clinicians can expect from a pulse oximeter; this time by adding the ability to monitor comfortably and reliably a patient's respiration rate. And, with In-Vivo Adjustment, clinicians can, when they need to, get very accurate measurements that are not only patient focused, but adjusted to their laboratory reference devices."

According to Michael O'Reilly, MD, Chief Medical Officer at Masimo, "On hospital general care floors around the world, the problems associated with aggressive post-operative pain management with patient-controlled analgesia (PCA) pumps, rising patient co-morbidities, and decreasing nurse-to-patient ratios have led to dramatic increases in adverse events and avoidable patients deaths. As a result, leading patient safety organizations have been calling for continuous monitoring of oxygenation and ventilation on post-surgical general care floor patients. Today, thanks to Masimo innovations, the need for a reliable SpO2 and respiration rate monitor for patients on general care floors is finally realized--and patients can be safer than ever before."

*The new Radical-7 Pulse CO-Oximeter has FDA 510(k) and CE Mark clearances for rainbow® Acoustic Monitoring and CE Mark clearance for In-Vivo Adjustment. In-Vivo Adjustment is pending FDA 510(k) clearance.

Source: Masimo Corporation

This collection is empty

Contact Us

Reviews

What our customers are saying

C. Lopez

Arrived very quickly. It was just what I was looking for and with a more approachable value.

Norman F

Great timely service, will use again.

Mark F

Great replacement battery!, it even has an expiration on it so that I know when to change it in 5 years. Last ones I purchased from another seller was garbage.

A.C.

Good turn around time

S.T.

Great product and service

R Harrison

Everything is as expected.